A study to examine the accuracy of potassium measurements in clinical laboratories across Canada
1Department of Pharmaceutical Sciences, University of Utrecht, Utrecht, The Netherlands.
Objective:
To assess the accuracy of potassium measurements in clinical laboratories across Canada.
Design And Method:
The flame atomic emission spectrophotometry reference method for the determination of potassium was established at the Canadian Reference Laboratory by using National Institute of Science and Technology standard reference materials. The method was subsequently used to assign target values for potassium to Canadian Reference Laboratory's External Quality Assessment human-serum-based testing material. A total of 503 laboratories participated and 9,279 individual External Quality Assessment test results were included in the study. Bias was determined by using difference plots.
Results:
Clinically significant bias (>1.6%) was observed in 45.9% of the laboratories. Bias ranged from 0.34 mmol/L to -0.54 mmol/L. At low concentrations (<3.5 mmol/L) a positive bias was most frequently observed (14.7% of analytical systems). At high potassium concentrations (>5.1 mmol/L) a negative bias was most frequently observed (31.4% of analytical systems).
Conclusion:
Inaccuracy in potassium results can contribute to test redundancy and mismanagement of patients, while prohibiting the merger of laboratory data from disparate testing sites for the purpose of trending and consolidation within a "universal health record." Inaccurate test results and the lack of standardization among laboratories adversely impact our ability to establish common reference intervals and critical limits. This inability has an adverse effect on medical decisions and patient care.
More Related Videos
06:24Generic Protocol for Optimization of Heterologous Protein Production Using Automated Microbioreactor Technology
Published on: December 15, 2017
10:38Observational Study Protocol for Repeated Clinical Examination and Critical Care Ultrasonography Within the Simple Intensive Care Studies
Published on: January 16, 2019
Related Concept Videos
Flame Photometry: Lab
Acute Kidney Injury IV: Diagnostic Studies and Prevention
Estimation of k and VD of Aminoglycosides