Water accumulation in metered dose inhaler spacers under normal mechanical ventilation conditions
Objective:
The purpose of this study was to compare the water accumulation in 3 types of metered dose inhaler (MDI) spacer shapes in-line in a ventilator circuit, in 2 positions over 2-, 4-, and 6-hour time periods through the use of heated- and nonheated-wire ventilator circuits.
Design:
The study design was prospective, quasiexperimental, and random assignment.
Setting:
The study was conducted in a university laboratory.
Materials:
Three brands of MDI spacers (OptiVent, ACE, AeroVent) were tested.
Outcome Measures:
Grams of water accumulation were measured.
Intervention:
Distilled water accumulation was measured in 3 brands of MDI spacers in 0 degrees and 45 degrees positions at 2-, 4-, and 6-hour time intervals. Water accumulation was measured in each spacer by calculating the differences between pretest (dry) weights and posttest (wet) weights through the use of an analytical balance. A Marquest SCT-3000 servo-controlled humidifier with heated-wire ventilator circuit was used with a room temperature range of 21.7 degrees C-22.8 degrees C (71 degrees -73 degrees F) and a relative humidity range of 57%-65%.
Results:
Multivariate repeated measures analysis demonstrated a difference between brands (P <.001). The amount of water accumulated during 6 hours (time variable) was significantly different (P <.001), as was the interaction between time and "spacer brand" (with Greenhouse-Geisser adjustment). The interaction of time and position was also significantly different (P =.001). Water accumulations at a 45 degrees angle were: AeroVent 0.765 +/- 0.152 g; OptiVent 1.894 +/- 0.228 g; and ACE 4.043 +/- 0.665 g through 6 hours of use.
Conclusions:
We found that water accumulation was a result of the type of spacer, position of the spacer, and time that the spacer was left in-line. All 3 brands of spacer collected less than 5 mL of water over 6 hours in either position. Heated-wire circuits accumulated less water than nonheated-wire circuits and may be safer when using MDI spacers.
Related Concept Videos
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
Pulmonary Ventilation: Inhalation
Boyle's law becomes particularly pertinent when examining respiratory...
Inhaled Medications
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...