A design for a dual assembly multileaf collimator.

P B Greer1, T van Doorn

  • 1Department of Medical Physics, Royal Adelaide Hospital, University of Adelaide, Australia. pgreer@mail.rah.sa.gov.au

Medical Physics
|December 1, 2000
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Remote EPID-based dosimetric auditing using DVH patient dose analysis.

Physics in medicine and biology·2023
Same author

A high-performance method of deep learning for prostate MR-only radiotherapy planning using an optimized Pix2Pix architecture.

Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB)·2022
Same author

Correlations between contouring similarity metrics and simulated treatment outcome for prostate radiotherapy.

Physics in medicine and biology·2018
Same author

Motion prediction in MRI-guided radiotherapy based on interleaved orthogonal cine-MRI.

Physics in medicine and biology·2016
Same author

Calculating radiotherapy margins based on Bayesian modelling of patient specific random errors.

Physics in medicine and biology·2015
Same author

MR simulation for prostate radiation therapy: effect of coil mounting position on image quality.

The British journal of radiology·2014
Same journal

Correction to "On the shape of the radiation survival curve in tumor spheroids: The role of oxygen heterogeneity".

Medical physics·2026
Same journal

Multi-view constrained semi-supervised vertebra detection for 3D ultrasound spine volume.

Medical physics·2026
Same journal

Accuracy of quantitative <sup>177</sup>Lu SPECT/CT imaging: A systematic review.

Medical physics·2026
Same journal

Physics-constrained dual-domain network for CBCT reconstruction from orthogonal X-rays in gynecologic radiotherapy.

Medical physics·2026
Same journal

Decomposition-based harmonization for quantitative PET imaging across scanners and radiotracers.

Medical physics·2026
Same journal

Development and evaluation of an in vivo dose-based monitoring system for electron FLASH radiation therapy.

Medical physics·2026
See all related articles

A novel multileaf collimator design enables rapid, grid-pattern imaging for radiation therapy. This innovation allows for faster double-exposure portal images, potentially acquired during treatment delivery.

Area of Science:

  • Medical Physics
  • Radiation Oncology
  • Imaging Technology

Background:

  • Current portal imaging methods in radiation therapy can be time-consuming.
  • Efficient imaging during treatment delivery is desirable for adaptive radiotherapy.

Purpose of the Study:

  • To investigate a novel multileaf collimator (MLC) design for rapid portal imaging.
  • To evaluate the imaging characteristics and shielding capabilities of the proposed MLC.

Main Methods:

  • A new MLC design featuring vertically displaced, laterally movable leaf banks was developed.
  • Radiation profiles were measured using tungsten and lead alloy collimators with varying parameters.
  • Image formation characteristics were analyzed based on slit width and collimator geometry.

Related Experiment Videos

Main Results:

  • The MLC design enables transmission of radiation in a grid pattern for imaging.
  • Slit width significantly influences transmitted profiles, with smaller widths causing broader profiles and overlap.
  • Adequate shielding was confirmed when adjacent leaf faces are aligned.

Conclusions:

  • The proposed MLC design facilitates rapid acquisition of double-exposure portal images.
  • The imaging characteristics are dependent on slit width, affecting image modulation or grid formation.
  • This technology has the potential to enhance imaging efficiency during radiation therapy delivery.