Related Experiment Videos
Variation in output factor caused by secondary blocking for 7-16 MeV electron beams
Medical Physics
|March 1, 1979
Abstract:
The effect of secondary blocking on the output for 7-16 MeV electron beams is investigated for the two most widely used methods of field shaping, i.e.,(a) shield on surface, and (b) shield at end of cone with subsequent air gap. For those two methods differences in output on the order of 10% are observed. These effects are explained in terms of the blocking geometry and its influence on scattered radiation.