Screening of Axonal Degeneration in Carpal Tunnel Syndrome Using Ultrasonography and Nerve Conduction Studies
Published on: January 11, 2019
Durkan gauge and carpal compression test: accuracy and diagnostic test properties
R S Wainner1, M L Boninger, G Balu
1Wilford Hall, USAF Medical Center, Lackland AFB, Texas, USA. rwainner@msn.com
The Durkan CTS gauge accurately measures pressure, but the carpal compression test (CCT) using this gauge is not a reliable diagnostic tool for carpal tunnel syndrome (CTS). Further research is needed for accurate CTS diagnosis.
Area of Science:
- Orthopedics
- Neurology
- Diagnostic Medicine
Background:
- The carpal compression test (CCT) is a diagnostic method for carpal tunnel syndrome (CTS).
- Previous studies suggest high sensitivity and specificity for CCT using thumb pressure.
- The diagnostic accuracy of the Durkan CTS gauge has not been independently validated.
Purpose of the Study:
- To evaluate the diagnostic performance of the CCT with the Durkan CTS gauge.
- To determine the measurement validity of the Durkan CTS gauge for assessing carpal tunnel pressure.
Main Methods:
- A prospective validity study was conducted.
- The Durkan CTS gauge's accuracy was verified using a force dynamometer.
- Nerve conduction studies (NCS) and CCT were performed on 52 participants (33 women, 19 men).
- CTS diagnosis was confirmed by compatible history and NCS results.
Main Results:
- The Durkan CTS gauge demonstrated accurate pressure readings.
- The CCT, when performed with the Durkan gauge, showed low sensitivity (.36) and specificity (.57) for CTS diagnosis.
Conclusions:
- The Durkan CTS gauge provides accurate pressure measurements.
- The CCT, utilizing the Durkan gauge, is not a sensitive or specific diagnostic test for carpal tunnel syndrome.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of the Cardiovascular System III: Palpation
Jugular Venous Pressure (JVP) Measurement
Position the patient at a thirty- to forty-five-degree angle or in a semi-fowler's position. Look for the highest point of pulsation in the internal jugular vein and measure the vertical distance to the angle of Loius or sternal angle. A normal JVP is 3-4 cm above the...
Peripheral Arterial Disease II: Clinical Manifestations and Diagnostic Evaluation