Durkan gauge and carpal compression test: accuracy and diagnostic test properties
R S Wainner1, M L Boninger, G Balu
1Wilford Hall, USAF Medical Center, Lackland AFB, Texas, USA. rwainner@msn.com
Study Design:
A prospective, criterion-based validity study.
Objectives:
To assess the diagnostic properties of the carpal compression test (CCT) when performed with the Durkan carpal tunnel syndrome (CTS) gauge, and to determine the measurement validity of the gauge.
Background:
The CCT has been reported to be highly sensitive (.87-.89) and specific (.93-1.0) in the diagnosis of CTS when it is done with thumb pressure. The accuracy of measurements with the Durkan CTS gauge, however, has not been established and the diagnostic sensitivity and specificity of the CCT when the gauge is used has not been independently confirmed.
Methods And Measures:
The study sample included 33 women and 19 men, aged 18 to 85 years (45.7 +/- 13.5 years). The accuracy of the gauge was determined with a force dynamometer and holding frame. Standard nerve conduction studies (NCS) and the CCT were performed on the symptomatic extremity of all subjects. A compatible history and the NCS results were used to confirm CTS.
Results:
The Durkan gauge registered pressures of 11.94 psi and 15.25 psi at the 12 and 15 psi gauge marks, respectively. Test sensitivity and specificity were .36 (95% CI = .17-.54) and .57 (95% CI = .39-.74), respectively.
Conclusions:
Pressure measurements obtained with the Durkan CTS gauge were accurate. The CCT when performed with the Durkan gauge, however, was neither sensitive or specific for the diagnosis of CTS.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of the Cardiovascular System III: Palpation
Jugular Venous Pressure (JVP) Measurement
Position the patient at a thirty- to forty-five-degree angle or in a semi-fowler's position. Look for the highest point of pulsation in the internal jugular vein and measure the vertical distance to the angle of Loius or sternal angle. A normal JVP is 3-4 cm above the...
Peripheral Arterial Disease II: Clinical Manifestations and Diagnostic Evaluation