Related Experiment Video
Updated: Aug 15, 2026

Surgical Placement of Catheters for Long-term Cardiovascular Exercise Testing in Swine
Published on: February 9, 2016
Mixed venous oxygen saturation as a promising parameter for physiologic control of total artificial heart
M Nakamura1, A Homma, E Tatsumi
1Department of Artificial Organs, National Cardiovascular Center Research Institute, Osaka, Japan.
Abstract:
Mixed venous oxygen saturation (SvO2) has been proposed as one of the suitable parameters for physiologic control of a total artificial heart (TAH). To establish the practical application of SvO2, we investigated the response of cardiac output (CO) and SvO2 to step-loaded exercise. A normal calf was surgically equipped with an ultrasonic flowmeter probe and an oximetry catheter in the pulmonary artery to measure CO and SvO2, respectively. Three stage step treadmill exercise tests (1, 2, and 4 km/h) were performed three times. While CO increased from 8.9 L/min at preexercise level to 9.7, 10.2 and 11.4 L/min at 1, 2, and 4 km/h, respectively, SvO2 decreased from 59.6% to 56.8, 55.3, and 52.2%, respectively. There existed a linear correlation between the magnitude of changes in CO and SvO2. CO and SvO2 exhibited a similar course of change, expressing an inverted exponential curve. The time constant of SvO2 was from 19 to 35 seconds, whereas that of CO was from 21 to 39 seconds. We conclude that SvO2 changes in close association with CO during exercise and has good potential to be a parameter for physiologic control of a TAH, by reflecting the recipient's CO demand without conspicuous time delay.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
03:40Point-of-Care Ultrasound for Peripheral Veno-Arterial Extracorporeal Membrane Oxygenation Without Left Ventricular Venting
Published on: January 17, 2025
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Physiological Control of Respiration
Breathing, a seemingly passive process, is regulated by the respiratory center in the brainstem. This center coordinates the involuntary control of respirations, which means it occurs without conscious effort, ensuring a smooth and uninterrupted pattern.
Regulation of Ventilation
The body maintains ventilation by monitoring levels of carbon dioxide (CO2), oxygen (O2), and hydrogen ion concentration (pH) in the arterial blood. Among these factors, the level of CO2 plays a crucial...
Venous Return
What is Venous Return?
Venous return refers to the rate at which blood flows back to the heart from the body's peripheral veins. It's an integral part of the circulatory system as it...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...