Continuous flow displacement immunosensors: a computational study
Related Experiment Video
Updated: Oct 10, 2026

Computer Numerical Control Micromilling of a Microfluidic Acrylic Device with a Staggered Restriction for Magnetic Nanoparticle-Based Immunoassays
Published on: June 23, 2022
D B Holt1, A W Kusterbeck, F S Ligler
1Naval Research Laboratory, Washington, DC 20375-5348, USA.
Abstract:
Numerical modeling has been used to investigate the disparity in performance and sensitivity that has been reported for flow displacement immunosensors based on bead-packed columns, membranes, and capillary tubes. The simulations strongly suggest that the high surface areas in the porous media systems may actually be detrimental to sensor performance because of large numbers of free antibody binding sites. Since the free antibody sites are created during the wash step in which the baseline is established, wash protocols are critical in optimizing the sensitivity for a given displacement sensor.