[Possible errors in leukocyte count caused by carryover in the Coulter counter S]
Abstract:
In 163 samples of venous blood with significant leukocytosis (20 000 to 172000 cells pro mm-3) a carryover of 400 to 22000 leukocytes up to the third count of the following specimens (200 to 9300 cells/mm-3) could be observed. Even after 41 specimens of normal leukocyte range and low leukocytosis (6700 to 17700 cells/mm-3) in cases of extreme leukopenias (less than 2000 cells/mm-3) the correct result was exceeded by 500 to 1000 cells. In four additional machines of other laboratories a carryover of different extent could be reproduced. In comparison to samples of venous blood only a very small carryover into subsequent samples of isotonic solution was found. The considerable spreading of the coefficient of variation between 1.7 and 16.2% (arithmetic mean 7.1%) prohibited a mathematical correction of error. Erroneously too high numbers of leukocytes could be absolutely avoided by at least two repeated counts.
More Related Videos
Related Concept Videos
Flow Cytometry
In...
Disorders of Leukocytes
Leukopenia may result from bone marrow disorders, autoimmune diseases, and infectious diseases. For example, conditions such as multiple myeloma and aplastic anemia can impair the bone marrow's ability to produce adequate leukocytes. Similarly, autoimmune diseases like lupus and viral infections such as HIV can prompt the immune system...