Related Experiment Video
Updated: Sep 4, 2026

Cardiopulmonary Bypass in a Mouse Model: A Novel Approach
Published on: September 22, 2017
Experimental use of a compact centrifugal pump and membrane oxygenator as a cardiopulmonary support system
1Department of Thoracic and Cardiovascular Surgery, Saga Medical School, Saga, Japan.
Abstract:
Compactness and high performance are the most important requirements for a cardiopulmonary support system. The Nikkiso (HPM-15) centrifugal pump is the smallest (priming volume; 25 ml, impeller diameter; 50 mm) in clinically available centrifugal pumps. The Kuraray Menox (AL-2000) membrane oxygenator, made of double-layer polyolefin hollow fiber, has a minimum priming volume (80 ml) and a low pressure loss (65 mm Hg at 2.0 L/min of blood flow) compared with other oxygenators. The aim of this study was to evaluate the performance of the most compact cardiopulmonary support system (total priming volume: 125 ml) in animal experiments. The cardiopulmonary bypass was constructed in a canine model with the Nikkiso pump and Menox oxygenator in comparison with a conventional cardiopulmonary support system. The partial cardiopulmonary bypass was performed for 4 h to evaluate the gas exchange ability, blood trauma, serum leakage, hemodynamics, and blood coagulative parameters. The postoperative plasma free hemoglobin level of the compact cardiopulmonary system was 29.5 +/- 10.21 mg/dl (mean +/- SD), which was lower than that of the conventional cardiopulmonary system, 48.75 +/- 27.39 mg/dl (mean +/- SD). This compact cardiopulmonary system provided the advantage in terms of reduction of the priming volume and less blood damage. These results suggested the possibility of miniaturization for the cardiopulmonary bypass support system in open-heart surgery in the near future.
Related Concept Videos
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...