Related Experiment Video
Updated: Aug 9, 2026

Removal of Trace Elements by Cupric Oxide Nanoparticles from Uranium In Situ Recovery Bleed Water and Its Effect on Cell Viability
Published on: June 21, 2015
Optimal sample preparation conditions for the determination of uranium in biological samples by kinetic
J W Ejnik1, M M Hamilton, P R Adams
1Applied Cellular Radiobiology Department, Armed Forces Radiobiology Research Institute, Bethesda, MD 20889-5603, USA. ejnik@mx.afrri.usuhs.mil
Abstract:
Kinetic phosphorescence analysis (KPA) is a proven technique for rapid, precise, and accurate determination of uranium in aqueous solutions. Uranium analysis of biological samples require dry-ashing in a muffle furnace between 400 and 600 degrees C followed by wet-ashing with concentrated nitric acid and hydrogen peroxide to digest the organic component in the sample that interferes with uranium determination by KPA. The optimal dry-ashing temperature was determined to be 450 degrees C. At dry-ashing temperatures greater than 450 degrees C, uranium loss was attributed to vaporization. High temperatures also caused increased background values that were attributed to uranium leaching from the glass vials. Dry-ashing temperatures less than 450 degrees C result in the samples needing additional wet-ashing steps. The recovery of uranium in urine samples was 99.2+/-4.02% between spiked concentrations of 1.98-1980 ng (0.198-198 microg l(-1)) uranium, whereas the recovery in whole blood was 89.9+/-7.33% between the same spiked concentrations. The limit of quantification in which uranium in urine and blood could be accurately measured above the background was determined to be 0.05 and 0.6 microg l(-1), respectively.