Failure of Cleaning Verification in Pharmaceutical Industry Due to Uncleanliness of Stainless Steel Surface
Published on: August 11, 2017
Evaluation of the effectiveness of decontamination of dental syringes
K Vickery1, A Pajkos, Y Cossart
1The Australian Centre for Hepatitis Virology and the Department of Infectious Diseases, University of Sydney, Australia.
Aim:
Steam autoclaving is the gold standard for decontaminating dental instruments, but worldwide disinfection is still widely employed. We have evaluated a range of procedures for their ability to inactivate duck hepatitis B virus contaminating dental syringes.
Methods:
Residual infectivity of virus suspensions following 2% glutaraldehyde treatment, ultrasonication or steam sterilisation at 121 degrees or 134 degrees was assayed by injecting day-old ducklings and examining their livers for viral DNA 2.5 weeks later. Dental syringes were contaminated with DHBV positive blood, then treated by the same methods. An anaesthetic cartridge containing water was loaded into the syringe and 400microl aliquots used to inject day-old ducklings. Used dental syringes were examined by Scanning Electron Microscopy.
Results:
Suspension test:- ultrasonic treatment failed to inactivate DHBV in suspension, but complete inactivation was achieved by 2% glutaraldehyde and autoclaving. Syringe test:- neither ultrasonic treatment nor glutaraldehyde inactivated DHBV. Autoclaving at 134 degrees (3 minutes) permitted transmission to 1/16 ducklings but steam sterilisation at 121 degrees (15 minutes) was effective. Electronmicroscopy demonstrated organic debris (biofilm) in the lumen of used syringes.
Conclusion:
Short autoclaving cycles, albeit at raised temperatures, may fail to inactivate the virus because of poor steam penetration, inadequate heat transfer and the accumulation of protective biofilm.
More Related Videos
Related Concept Videos
Hand hygiene
Hand washing...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Antimicrobial Effectiveness