Skin disease presenting as an outbreak of pseudobacteremia in a laboratory worker
1Department of Clinical Microbiology and Infectious Diseases, Hadassah University Hospital, Ein Kerem, Jerusalem. simhon@hadassah.org.il
Abstract:
An outbreak of pseudobacteremia due to Streptococcus pyogenes (group A streptococci [GAS]) and methicillin-susceptible Staphylococcus aureus (MSSA) was traced to the venting procedure for aerobic bottles prior to their loading into the incubator of the BacT/Alert analyzer (Organon Teknika). Bacteria shed by a laboratory worker suffering from impetigo and cellulitis contaminated the aerobic bottles of 10 patients. All blood culture isolates, in addition to the isolates from the laboratory worker, were of the same GAS M and T types. All MSSA isolates from blood cultures and the index case's hands had the same lytic phage profile. Procedural breakdowns were identified in the laboratory. Bottles were vented outside the biological safety cabinet, gloves were not worn, and unprotected needles were used for the venting procedure. The source of the aspirated bacteria that contaminated the bottles was identified and the index case was treated promptly.
Insights
A laboratory worker
Area of Science:
- Clinical Microbiology
- Infectious Disease Epidemiology
Background:
- Pseudobacteremia outbreaks pose a diagnostic challenge.
- Contamination of blood cultures can lead to misdiagnosis and inappropriate treatment.
Purpose of the Study:
- To identify the source and cause of a pseudobacteremia outbreak.
- To evaluate laboratory procedures contributing to contamination.
Main Methods:
- Investigated an outbreak of bloodstream infections.
- Performed bacterial isolate typing (M and T types for GAS, phage profiling for MSSA).
- Reviewed laboratory protocols for blood culture handling.
Main Results:
- Identified Streptococcus pyogenes (group A streptococci [GAS]) and methicillin-susceptible Staphylococcus aureus (MSSA) as causative agents.
- Traced contamination to improper venting of aerobic blood culture bottles by an infected laboratory worker.
- Identified procedural breaches including inadequate personal protective equipment and unsafe venting techniques.
Conclusions:
- Improper laboratory procedures, specifically bottle venting, led to patient contamination.
- Strict adherence to aseptic techniques and biosafety protocols is crucial to prevent pseudobacteremia.
- Prompt identification and treatment of the source case resolved the outbreak.
Related Concept Videos
Skin Diseases and Disorders
Gram-positive Staphylococcus spp. and Streptococcus spp. are responsible for many of the most common skin infections. However, many...
Defense Against Bacterial Pathogens
Phagocytes
Phagocytes are the frontline soldiers of the immune system. They include neutrophils and macrophages. Neutrophils are the most abundant type of white blood cell and are quickly mobilized to the site of infection. Macrophages are larger cells that patrol...
Investigation of Disease Outbreaks
Staphylococcal Skin Infections
Atypical Pneumonia