Central venous pressure measurements: peripherally inserted catheters versus centrally inserted catheters
I H Black1, S A Blosser, W B Murray
1Pennsylvania State University College of Medicine, Hershey 17033-0850, USA.
Objective:
To determine whether central venous pressure measurements taken from a peripherally inserted central catheter (PICC) correlate with those from a centrally inserted central catheter (CICC).
Design:
A pilot bench study followed by a prospective, non-blinded, clinical comparison.
Setting:
A 16-bed medical coronary intensive care unit and a 30-bed surgical intensive care unit at a university hospital.
Patients:
Seven surgical intensive care unit patients and five medical coronary intensive care unit patients.
Interventions:
During the bench study, a simple manometer system was set up to test the catheters. During the clinical study, measurements of central venous pressure were recorded from patients who had an indwelling CICC and PICC concomitantly. Positions of the catheter tips in the chest were verified by radiography. Paired central venous pressure measurements were taken from 19-gauge dual-lumen PICCs and from 7-Fr, 16-gauge, 18-gauge, and pulmonary artery catheter CICCs, all with continuous pressure infusion devices.
Measurements And Main Results:
Bench work showed that PICCs, because of their longer length and narrower lumen, have a higher inherent resistance, which can be overcome with a continuous infusion device. During the clinical study, three to 12 paired, digital, central venous pressure measurements were recorded from each of 12 patients for a total of 77 data pairs. Measurements were recorded at end-expiration. Mean central venous pressure from the CICCs was 11 + 7 mm Hg, and from the PICCs was 12 + 7 mm Hg. PICC pressure versus CICC pressure correlated (r = 0.99) for all data pairs. Analysis by repeated measures showed PICC central venous pressure more than CICC central venous pressure by 1.0 + 3.2 mm Hg (p = 0.02).
Conclusions:
PICCs can be used to measure central venous pressure and to follow trends in a clinical setting when used with a pressure infusion device to overcome the natural resistance of the PICC. Central venous pressure recorded via PICCs is slightly higher, but the difference is clinically insignificant.
More Related Videos
09:57A Protocol to Set Up Needle-Free Connector with Positive Displacement on Central Venous Catheter in Intensive Care Unit
Published on: July 13, 2019
05:51Computed Tomography (CT) Guided Implantation of a Totally Implantable Venous Access Port (TIVAP) through Subclavian Vein
Published on: January 13, 2026
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pre-Procedural Guidelines for Assessing Blood Pressure
Parentral Nutrition: Centeral and Peripheral Parental Nutrition
PN can be administered through two primary routes:
1. Central Parenteral Nutrition (CPN):
CPN involves delivering a high concentration of nutrients through a large vein. This is typically achieved using a Peripherally Inserted Central Catheter (PICC) or,...
Cardiac Catheterization II: Right Heart Catheterization
Cardiac Catheterization III: Left Heart Catheterization