The Supraclavicular Fossa Ultrasound View for Central Venous Catheter Placement and Catheter Change Over Guidewire
Published on: December 23, 2014
Central venous pressure measurements: peripherally inserted catheters versus centrally inserted catheters
I H Black1, S A Blosser, W B Murray
1Pennsylvania State University College of Medicine, Hershey 17033-0850, USA.
Peripherally inserted central catheters (PICCs) can accurately measure central venous pressure, correlating well with centrally inserted central catheters (CICCs). Using a pressure infusion device with PICCs ensures reliable central venous pressure monitoring in clinical settings.
Area of Science:
- Cardiovascular Medicine
- Critical Care Medicine
- Medical Devices
Background:
- Central venous pressure (CVP) monitoring is crucial in intensive care.
- Traditionally, CVP is measured using centrally inserted central catheters (CICCs).
- Peripherally inserted central catheters (PICCs) offer a less invasive alternative but their utility for CVP measurement requires validation.
Purpose of the Study:
- To evaluate the correlation between CVP measurements obtained from PICCs and CICCs.
- To determine the clinical applicability of PICCs for CVP monitoring.
Main Methods:
- A pilot bench study assessed catheter resistance.
- A prospective, non-blinded clinical study compared CVP measurements from concomitant PICCs and CICCs in 12 ICU patients.
- Measurements were taken using digital systems with continuous pressure infusion devices.
Main Results:
- Bench study indicated PICCs have higher resistance, manageable with infusion devices.
- Clinical study yielded 77 paired CVP measurements (mean CVP: CICC 11±7 mm Hg, PICC 12±7 mm Hg).
- A strong correlation (r=0.99) was observed between PICC and CICC CVP readings; PICC CVP was slightly higher (1.0±3.2 mm Hg, p=0.02).
Conclusions:
- PICCs are suitable for CVP measurement and trend monitoring in clinical settings.
- Utilizing a pressure infusion device is essential to overcome PICC's inherent resistance.
- The slight increase in CVP measured by PICCs is clinically insignificant, validating their use.
More Related Videos
09:57A Protocol to Set Up Needle-Free Connector with Positive Displacement on Central Venous Catheter in Intensive Care Unit
Published on: July 13, 2019
05:51Computed Tomography (CT) Guided Implantation of a Totally Implantable Venous Access Port (TIVAP) through Subclavian Vein
Published on: January 13, 2026
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pre-Procedural Guidelines for Assessing Blood Pressure
Parentral Nutrition: Centeral and Peripheral Parental Nutrition
PN can be administered through two primary routes:
1. Central Parenteral Nutrition (CPN):
CPN involves delivering a high concentration of nutrients through a large vein. This is typically achieved using a Peripherally Inserted Central Catheter (PICC) or,...
Cardiac Catheterization II: Right Heart Catheterization
Cardiac Catheterization III: Left Heart Catheterization