Dosimetry for Cell Irradiation using Orthovoltage (40-300 kV) X-Ray Facilities
Published on: February 20, 2021
Dosimetry of rotational partial-skin electron irradiation
K Müller-Sievers1, E Ertan, B Kober
1Klinikum Darmstadt, Department of Radiooncology/Medical Physics, Grafenstrasse 9, D-64283 Darmstadt, Federal Republic of Germany.
Background And Purpose:
Often, the most appropriate treatment for superficially and extensively spreading tumors of the skin is to use electron irradiation at enlarged distances. Rotational skin electron irradiation is a proven method for the treatment of the entire skin surface. We here report modifications of this technique in the set-up of partial-skin electron irradiation and the results of dosimetric examinations with regard to optimal shielding, dose profiles and depth dose curves under various irradiation conditions.
Materials And Methods:
Irradiation was performed using electron beams with nominal energies of 6 MeV from a linear accelerator. The phantom was located on a rotating platform at a source-surface distance SSD=300 cm. A horizontal slit aperture (height: 32 cm) within a 2 cm thick polymethylmethacrylate (PMMA) shielding plate near the phantom was used to define the size of the irradiated region. Influences on dose distributions due to scattering processes on the PMMA edges were investigated using a flat ionization chamber and films. Absolute dose measurements and film calibration were made with the flat chamber. The quality of bremsstrahlung radiation behind the shielding was determined with a thimble ionization chamber in the phantom.
Results And Conclusions:
The results of rotational partial-skin electron irradiation reveal some of the investigated shielding geometries to be optimal. Depth dose distributions and dose rates correspond to the results obtained in total skin electron rotational irradiation. It is possible to apply the dose superficially in the first millimeters of the skin; the dose maximum is located at a depth of 0-2 mm, the 80% isodose at 9 mm. The amount of bremsstrahlung contamination is 2.5%. The local amount of absorbed dose per monitor unit depends strongly on patient/phantom cross-section geometry. At our institute, rotational partial-skin electron irradiation was implemented into clinical routine in 1997.
More Related Videos
06:21Diffuse Optical Spectroscopy for the Quantitative Assessment of Acute Ionizing Radiation Induced Skin Toxicity Using a Mouse Model
Published on: May 27, 2016
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
Biological Effects of Radiation
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...