Comparative performance of impactor air samplers for quantification of fungal contamination

D Nesa1, J Lortholary, A Bouakline

  • 1Laboratoire de Parasitologie-Mycologie, Hôpital Saint-Antoine, France.

Insights

Four impactor air samplers were evaluated for fungal spore collection in hospitals. Air Test Omega, Mas-100, and BioImpactor showed comparable performance, making them suitable for routine hospital air monitoring.

Area of Science:

  • Environmental Microbiology
  • Hospital Infection Control

Background:

  • Fungal spore contamination in hospitals poses a risk to patient health.
  • Accurate air sampling is crucial for assessing and managing indoor fungal bioburden.

Purpose of the Study:

  • To evaluate the performance of four impactor air samplers for fungal spore collection.
  • To identify suitable air sampling devices for routine hospital indoor air quality assessment.

Main Methods:

  • Comparison of four impactor air samplers (Samplair, Air Test Omega, Air Samplair Mas-100, BioImpactor 100-08) in three hospital environments.
  • Further comparison of the top-performing impactor (BioImpactor) against filtration and centrifugation-based samplers.

Main Results:

  • Air Test Omega, Mas-100, and BioImpactor demonstrated comparable and efficient fungal spore recovery.
  • Samplair showed significantly lower efficiency compared to the other impactors.
  • No significant difference in spore recovery was found between the superior impactors and filtration/centrifugation methods.

Conclusions:

  • Air Test Omega, Air Samplair Mas-100, and BioImpactor 100-08 are suitable for routine indoor fungal contamination evaluation in hospitals.
  • These samplers offer a balance of performance, cost, autonomy, and ease of use for clinical environments.