Home ventilator low-pressure alarms fail to detect accidental decannulation with pediatric tracheostomy tubes
S S Kun1, C T Nakamura, J F Ripka
1Division of Pediatric Pulmonology, Childrens Hospital Los Angeles, Keck School of Medicine of the University of Southern California, Los Angeles, CA 90027, USA.
Background:
Positive-pressure ventilators are equipped with low-inspiratory-pressure alarms to protect patients from hypoventilation. Small uncuffed tracheostomy tubes have a high resistance, and may not trigger these alarms during decannulation.
Study Objective:
To determine whether ventilator low-inspiratory-pressure alarms are effective in detecting tracheostomy decannulation.
Design:
We connected tracheostomy tubes of varying inner diameters (3.0 to 6.0 mm) to a home ventilator and simulated decannulation using low (tidal volume [VT], 600 mL; peak inspiratory pressure [PIP], 25 cm H(2)O), medium (VT, 800 mL; PIP, 30 cm H(2)O), and high (VT, 1,000 mL; PIP, 35 cm H(2)O) ventilator settings.
Results:
When the ventilator low-inspiratory-pressure alarm was set at 4 cm H(2)O below the desired PIP, it failed to alarm for simulated decannulation of tracheostomy tubes < 4.5 mm on low and medium settings, and < 4.0 mm on high settings. When the ventilator low-inspiratory-pressure alarm was set at 10 cm H(2)O below the desired PIP, it failed to alarm with tracheostomy tubes < 6.0 mm.
Conclusion:
We conclude that ventilator low-inspiratory-pressure alarms fail to alarm during simulated decannulation with small tracheostomy tubes commonly used in children. We speculate that low-inspiratory-pressure alarms set at 4 cm H(2)O below the desired PIP will detect more decannulation than when set at 10 cm H(2)O below the desired PIP.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...