High spatial resolution measurements of specific absorption rate around ICD leads
B B Beard1, M S Mirotznik, I A Chang
1Food and Drug Administration, Center for Devices and Radiological Health, Electrophysics Branch, HFZ-133, 9200 Corporate Boulevard, Rockville, MD 20850, USA. bbb@cdrh.fda.gov
Introduction:
It has been shown that strong electric shocks can cause local refractoriness in the heart. This is of particular concern if the region of refractoriness is the area sensed by an implant to determine cardiac rhythm, as is the case with many Implantable Cardioverter Defibrillator (ICD) leads which use the same electrodes for shocking and sensing. Failure to sense the true cardiac rhythm can cause application of unnecessary shocks and potential induction of arrhythmias. We developed a system to accurately map the areas where local refractoriness is most probable. We measured the Specific Absorption Rate (SAR) around typical ICD leads. Current density (J), a parameter that determines defibrillation effectiveness, is proportional to the square root of SAR.
Methods And Results:
SAR measurements were performed in a homogeneous saline media using a variety of ICD leads. Gated 60 Hz shocks were used to produce heating, which was measured by thermistor probes. The temperature-rate-of-change is directly proportional to the SAR. Measurement techniques were developed that produced accurate SAR results at high spatial resolutions. Multiple polarities and configurations of ICD leads were tested.
Conclusions:
We confirmed the spatial distribution of the SAR and corresponding current density possessed sharp peaks and were highly localized around the leads' electrodes. Scans with a resolution of 1 mm or less are required in the area of peak SAR in order to capture the peak's value.
More Related Videos
06:28Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
09:46Direct Comparison of Hyperspectral Stimulated Raman Scattering and Coherent Anti-Stokes Raman Scattering Microscopy for Chemical Imaging
Published on: April 28, 2022
Related Concept Videos
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
Atomic Emission Spectroscopy: Lab