Clinical impact of arrhythmias on gated SPECT cardiac myocardial perfusion and function assessment
1Division of Cardiology, Columbia University, PH9-993B, 622 W 168th St., New York, NY 10032, USA. kjn3@columbia.edu
Background:
We reported previously that mean quantified cardiac functional parameters computed by one gated single photon emission computed tomography (SPECT) technique were not significantly altered by common gating errors. However, it is not known to what extent other gated SPECT approaches that are based on different ventricular modeling assumptions are influenced by arrhythmias, nor are the effects of gating errors on visual analyses and their subsequent clinical implications known.
Methods:
Projection data for 50 patients (aged 64 +/- 12 years; 68% men; 76% with myocardial perfusion defects) undergoing technetium-99m sestamibi gated SPECT who were in sinus rhythm during data acquisition were altered to simulate common arrhythmias. To determine quantitative effects, we performed calculations for original control and altered images by Gaussian myocardial detection (Quantitative Gated SPECT [QGS] program) and by wall thickening derived from gated perfusion polar maps (Emory Cardiac Toolbox program). To evaluate visual assessment in control and simulated-arrhythmia tomograms, 2 experienced blinded observers independently interpreted perfusion from polar maps and wall motion and thickening from tomographic cines, using a 4-point scale.
Results:
Although mean functional parameters were scarcely altered, paired t tests showed ejection fraction fluctuations to be significantly different from control values, causing patients to change between abnormal and normal ejection fraction categories (2% of patients by QGS and 14% by Emory Cardiac Toolbox). Visual examination of QGS polar perfusion and function maps showed changes for 72% of cases, although in only 4% were these considered to have potential clinical consequences. The kappa statistic for visual analysis of concordance between control and arrhythmia readings showed that agreement was "excellent" for perfusion, "good" for motion, and "marginal" for thickening.
Conclusions:
As with quantitative measurements, thickening is the parameter most prone to error in the presence of arrhythmias. It is important to test data for gating errors to avoid potentially erroneous measurements and visual readings.
More Related Videos
09:10Normothermic Ex Situ Heart Perfusion in Working Mode: Assessment of Cardiac Function and Metabolism
Published on: January 12, 2019
08:51Intra-cardiac Side-Firing Light Catheter for Monitoring Cellular Metabolism using Transmural Absorbance Spectroscopy of Perfused Mammalian Hearts
Published on: May 12, 2019
Related Concept Videos
Mechanism of Cardiac Arrhythmias
Pathophysiology of Cardiac Performance
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and the T...
Dysrhythmias V: Evaluating Dysrhythmias