[Mid-term complications of automatic implantable cardiac defibrillators]
D Babuty1, L Fauchier, J C Charniot
1Service de chirurgie cardiaque, centre hospitalier universitaire Trousseau, 37044 Tours.
Insights
Automatic implantable cardiac defibrillators (AICD) effectively treat ventricular arrhythmias but carry significant medium-term complication risks. Careful specialist follow-up is crucial due to the incidence and variety of AICD-related issues.
Area of Science:
- Cardiology
- Medical Devices
- Electrophysiology
Context:
- Automatic implantable cardiac defibrillators (AICD) are vital for managing malignant ventricular arrhythmias.
- Patient populations often have underlying cardiac conditions and require long-term monitoring.
- Understanding medium-term complications is essential for optimizing patient care and device management.
Purpose:
- To investigate the incidence and nature of medium-term complications associated with AICD implantation.
- To analyze complication types across operative, postoperative, and medium-term follow-up periods.
- To evaluate the overall safety profile and management needs for AICD recipients.
Summary:
- A study of 79 AICD implants in 50 patients (average follow-up 41.3 months) revealed 42% experienced complications.
- Complications included operative issues (e.g., AV block, electrode rupture), postoperative events (e.g., infection, CVA), and medium-term problems (e.g., inappropriate shocks, device failures).
- Inappropriate shocks were common, often linked to atrial arrhythmias or lead dysfunction, alongside material-related issues like electrode rupture.
Impact:
- AICD therapy is effective for life-threatening arrhythmias but necessitates vigilant specialist follow-up.
- Findings highlight the need for proactive management strategies to mitigate AICD-related morbidity.
- This research informs clinical practice regarding patient selection, monitoring protocols, and complication management for AICD patients.
Abstract:
The incidence and the nature of medium-term complications of automatic implantable cardiac defibrillators (AICD) were studied. Seventy-nine AICD were implanted in 50 consecutive patients (42 men, aged 54.5 +/- 13.7 years). Forty-six patients had spontaneous ventricular arrhythmia. These arrhythmias were resistant to treatment (N = 9), reproducible with treatment (N = 28). In 4 patients, the indication was prophylactic, in 2 a Brugada syndrome, in 2 syncope with reinducible ventricular tachycardia and in 1 patient, torsades with a short coupling interval. Forty-six patients had underlying cardiac disease (ischaemic, N = 28, primary dilated cardiomyopathy, N = 10, others, N = 8). The ejection fraction was > 40% in 32 patients. The average follow-up was 41.3 +/- 34.9 months. Eight patients died, 2 from cardiac failure. Twenty-one patients (42%) had 1 or more complications related to their AICD. These occurred: in the operative period (N = 3): 1 post-shock atrioventricular block, 1 ruptured electrode and 1 increased threshold with amiodarone; in the postoperative period (N = 6): infection in 3 cases, cerebrovascular accident in 1 case, deep venous thrombosis of the left arm in 1 case, pneumothorax in 1 case. In the medium-term, the complications were mainly inappropriate electrical shocks observed in 14 patients related to atrial arrhythmias in 7 cases, sinus tachycardia in 1 case, over-detection of myopotentials in 2 cases and electrode dysfunction in 4 cases. In addition, the authors observed complications related to the material: AICD failure in 1 case, electrode displacement in 1 case, and electrode rupture in 3 cases. The authors conclude that AICD are effective for the treatment of malignant ventricular arrhythmias which justify strict specialist follow-up given the incidence and diversity of their complications.
Related Concept Videos
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Cardiac Catheterization I: Pre-Procedure Overview
Dysrhythmias VI: Management of Dysrhythmias
Cardiopulmonary Resuscitation III: AED Use
Cardiomyopathy II: Dilated Cardiomyopathy
Cardiomyopathy V: Interprofessional Care