[Problems in clinical use of intraarterial blood gas monitoring system, Paratrend 7]
1Department of Anesthesiology, Kanagawa Cardiac and Respiratory Center, Yokohama 236-0051.
Abstract:
We evaluated the intraarterial blood gas monitoring system Paratrend 7 in clinical use. One hundred and sixty six sensors used for patients receiving thoracic, open heart or great vessel surgery were tested for one year, prospectively. Sensors were inserted through a 20 G intraarterial cannula into the radial artery before operation and removed when they malfunctioned or seemed to be not necessary for patient management in critical care unit. We encountered several problems such as, initial calibration failure (5/166), impossible sensor insertion (2/166), arterial pressure wave change (5/166), parameter display failure (21/166), and thrombus in the circuit (9/166). Parameter display failure after insertion tended to occur on patient's arm movement imparing the sensor. No serious accident, such as arterial thrombosis was experienced. Some troubles were due to staff's incorrect handling of the apparatus. Training of staffs is indispensable for safe and sufficient utilization of Paratrend 7.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to reabsorb or...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...