Late complications of aortic valve replacement with cloth-covered, composite-seat prostheses. A six-year appraisal

Insights

Warfarin significantly reduces thromboembolic complications in patients with cloth-covered aortic prostheses. Antiplatelet drugs have not yet proven sufficient for preventing these adverse events.

Area of Science:

  • Cardiovascular Surgery
  • Biomaterials Science

Background:

  • Cloth-covered composite-seat aortic prostheses are widely used.
  • Assessing long-term safety and efficacy is crucial for patient outcomes.

Purpose of the Study:

  • To evaluate the early and late results of cloth-covered aortic prostheses.
  • To compare thromboembolic complication rates with and without warfarin.
  • To investigate the incidence of strut cloth wear.

Main Methods:

  • Advanced actuarial techniques applied to a cohort of 396 patients.
  • Analysis of postoperative warfarin use versus no warfarin.
  • Examination of explanted prostheses for wear at reoperation.

Main Results:

  • Warfarin use in 116 patients with Model 2310-2320 prostheses resulted in zero thromboembolic complications over 360 patient-years.
  • Patients with the same prosthesis but without warfarin experienced 9 emboli per 100 patient-years.
  • Cloth wear was observed in 10 patients, a complication addressed by the Model 2400 composite strut valve.

Conclusions:

  • Concomitant anticoagulant therapy, specifically warfarin, significantly enhances the safety of cloth-covered aortic valves.
  • The efficacy of antiplatelet agents alone has not been established for these prostheses.
  • The Model 2400 valve design mitigates strut cloth wear.