Related Experiment Video
Updated: Jul 16, 2026

A Saline/Bipolar Radiofrequency Energy Device As an Adjunct for Hemostasis in Solid Organ Injury/Trauma
Published on: July 28, 2020
Early experiences of haemostasis on brain tumour surgery with Argon Plasma Coagulation (APC)
T Miyazawa1, H Nawashiro, K Shima
1Department of Neurosurgery, National Defense Medical College, Tokorozawa, Saitama, Japan.
Objective:
We first applied a novel haemostatic strategy involving Argon Plasma Coagulation (APC), an innovative no-touch electrocoagulation technique in which a high-frequency alternating current is delivered to the tissue through ionized argon gas, to brain tumour surgery, and report its usefulness and limitations.
Methods:
The APC system we used comprised an APC 300 developed by ERBE Elektromedizin GmbH, Germany. We applied APC to 13 brain tumours in 12 patients (5 meningiomas, 4 sarcomas, 2 glioblastomas, and 2 pituitary adenomas). To avoid unnecessary thermal injury to the tissue as much as possible, power/gas flow settings of 20 and 40 W were used. The impact time was varied individually but was around several seconds per one impact. The argon jet (1.5-4.5 L/min) clears a field of pooled blood and evenly conducts electrical energy to the target tissue. A thin and flexible probe particularly increased the usefulness of APC for haemostasis on deep-seated skull base tumour operations under a microscope.
Conclusion:
All patients were successfully treated and satisfied with the surgical results without any complications due to APC. APC appears to be an excellent alternative strategy for achieving haemostasis on vascular-rich brain tumour surgery, and may be valuable for the management of patients with coagulation defects.