Characteristics and results of epicardial pacing in neonates and infants

E Villain1, H Martelli, D Bonnet

  • 1Service de Cardiologie Pédiatrique, Hôpital Necker Enfants-Malades, Paris, France.

Insights

Epicardial pacing effectively treats congenital complete atrioventricular block (CAVB) in infants. This study shows satisfactory outcomes with no pacing-related mortality and improved thresholds using steroid-eluting electrodes.

Area of Science:

  • Pediatric Cardiology
  • Cardiac Electrophysiology
  • Congenital Heart Disease

Background:

  • Congenital complete atrioventricular block (CAVB) in infants often requires pacing for symptomatic bradycardia or associated arrhythmias.
  • Traditional pacing methods present challenges in pediatric populations, including lead dislodgement and generator issues.

Purpose of the Study:

  • To evaluate the efficacy and safety of epicardial pacing in infants diagnosed with isolated CAVB.
  • To assess long-term outcomes, including pacing thresholds, generator longevity, and complication rates.

Main Methods:

  • Retrospective analysis of 34 infants (aged 1 day to 20 months) who underwent epicardial pacing between 1988 and 1998.
  • Implantation via left thoracotomy or subxyphoid approach, with a preference for dual-chamber units in 22 patients.
  • Use of steroid-eluting electrodes was noted to improve chronic ventricular thresholds.

Main Results:

  • No operative mortality; three late deaths attributed to cardiomyopathy.
  • Seven infants required reoperation for complications such as electrode displacement, infection, or exit block.
  • Chronic ventricular thresholds were generally low (0.3-2 V), with improved thresholds noted with steroid-eluting electrodes.
  • Pacemaker generators demonstrated longevity, lasting at least 5 years in 16 cases.

Conclusions:

  • Epicardial pacing is a safe and effective treatment for infants with CAVB, yielding satisfactory long-term results.
  • The deep generator placement minimizes risks of cutaneous erosion and allows for larger, longer-lasting devices.
  • Steroid-eluting electrodes contribute to improved pacing thresholds and potentially better long-term outcomes.

Related Concept Videos

Pulse rhythm01:30

Pulse rhythm

Pulse rhythm refers to the pattern of pulsations within specific intervals, offering valuable insights into the regularity or irregularity of the heart's beats as observed through the pattern of pulsation within specific intervals. A regular pulse exhibits a consistent heart rate with uniform waveforms and pulsation force, variations of which can be classified as normal, weak, or bounding.
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Electrophysiology of Normal Cardiac Rhythm01:19

Electrophysiology of Normal Cardiac Rhythm

The normal cardiac rhythm is a synchronized electrical activity that facilitates the regular and coordinated contraction of the heart muscle. This process is essential for efficient blood circulation throughout the body. The fundamental elements involved in establishing and maintaining this rhythm include the unique electrical properties of cardiac muscle cells, the sinoatrial (SA) node's pacemaker function, the specialized conducting system, and the ionic mechanisms underlying each phase of...
Cardiopulmonary Resuscitation III: AED Use01:23

Cardiopulmonary Resuscitation III: AED Use

Introduction to AEDAn Automated External Defibrillator (AED) is a portable medical device that analyzes the heart's rhythm and, if necessary, delivers an electrical shock to help the heart re-establish an effective rhythm during sudden cardiac arrest (SCA). SCA occurs when the heart suddenly and unexpectedly stops beating, leading to a loss of blood flow to the brain and other vital organs. In such emergencies, time is of the essence, and using an AED, combined with Cardiopulmonary...