Related Experiment Video
Updated: Jul 14, 2026

A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Lead noise with an active-fixation defibrillation lead.
R N Gelder1, J M Galvin, C M Albert
1Cardiac Arrhythmia Service, Massachusetts General Hospital, Fruit Street, Boston, MA 02114, USA. rgelder@partners.org
A specific implantable cardioverter defibrillator lead can cause lead artifact, leading to inappropriate detections. This unique issue often resolves within four weeks, preventing unnecessary therapies.
Area of Science:
- Cardiology
- Biomedical Engineering
Background:
- Inappropriate detection and therapy are common adverse events in implantable cardioverter defibrillator (ICD) therapy.
- Lead artifact, often presenting late due to lead component stress, is a known mechanism for these inappropriate events.
Observation:
- This study describes experience with Endotak Endurance EZ leads (Models 0154/0155/0156).
- Of 20 implants, four patients exhibited noise leading to inappropriate detections in three.
- No patient received therapy due to these detections.
Findings:
- The artifact appeared early post-implant and resolved in three of four affected patients by four weeks.
- No noise developed in the remaining 16 patients after this initial period.
- This specific lead problem is unique in its transient nature.
Implications:
- While resolving over time, this lead artifact can still cause inappropriate therapies and management challenges.
- Careful monitoring is needed, especially in high-risk patients with ventricular arrhythmias.
- Understanding lead behavior is crucial for optimizing ICD therapy safety and efficacy.