Related Experiment Video
Updated: Oct 9, 2026

Full-root Aortic Valve Replacement by Stentless Aortic Xenografts in Patients with Small Aortic Roots
Published on: May 21, 2017
Early clinical experience with the On-X prosthetic heart valve
A Laczkovics1, M Heidt, H Oelert
1Universitätsklinikum Bergmannsheil, Bochum, Germany.
Background And Aim Of The Study:
The study was designed to investigate the clinical performance of the On-X prosthetic heart valve in a multicenter experience.
Methods:
Between September 1996 and September 1999, 301 patients (56% males) underwent isolated On-X valve replacement (184 isolated aortic (AVR), 117 isolated mitral (MVR)) at 11 European centers under a standardized protocol. Average age at implant was 60.2 years. Office or hospital follow up was required by the protocol; average follow up on all patients was 11 months.
Results:
Thirty-day mortality in the study was 2.2% for AVR and 6.0% for MVR, with valve-related mortality of 0.5% for AVR. There were eight late deaths (0.7%/pt-yr AVR and 2.3%/pt-yr MVR). Two of these deaths were sudden, and thus possibly valve-related (one AVR, one MVR). Early total valve-related morbidity was 3.5% for AVR and 2.6% for MVR. In total, 13 thromboembolic events occurred; one early event in AVR resulted in death (0.5%), and one transient early event occurred in MVR (0.9%). There were 11 late events (seven AVR (1.7%/pt-yr) and four MVR (1.8%/pt-yr)), for a two-year freedom from thromboembolism of 96.6% after AVR and 97.1% after MVR. Three late bleeding events occurred, all after AVR (0.7%/pt-yr and 98.9% free at two years). Major paravalvular leaks requiring reoperation occurred on two occasions early (one AVR (0.5%), one MVR (0.8%)) and once late in MVR (0.5%/pt-yr). Late minor, untreated paravalvular leaks occurred in three AVR patients (0.7%/pt-yr) and in one MVR patient (0.5%/pt-yr). Prosthetic endocarditis occurred four times (two AVR (0.5%/pt-yr), two MVR (0.9%/pt-yr)), all within the first 12 months of surgery. Actuarial freedom from all valve-related events at two years was 88.7% for AVR and 88.1% for MVR. NYHA class was improved in 75.8% of AVR patients and 70.6% of MVR patients at two years after surgery.
Conclusion:
These early results indicate that the On-X valve provides satisfactory clinical outcome in the immediate postoperative period, and that the valve is both safe and effective.
Related Concept Videos
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Mitral Valve Prolapse I: Introduction
Mitral Stenosis I: Introduction
Mitral Regurgitation I: Introduction
Aortic Regurgitation I: Introduction
Mitral Stenosis II: Clinical features and Diagnostic Tests