A system to regulate aortic pressure in the rat.

D M Smith

    Journal of Applied Physiology
    |February 1, 1975
    PubMed
    Summary
    This summary is machine-generated.

    Researchers developed a servo-controlled aortic clamp for precise control of blood pressure in rats. This new device enables easier study of how arterial pressure changes affect organ function.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    Relative accuracy of algorithm-based prescription of nasal CPAP in OSA.

    Respiratory medicine·2004
    Same author

    A novel, simplified approach to starting nasal CPAP therapy in OSA.

    Respiratory medicine·2004
    Same author

    CL/RAMP2 and CL/RAMP3 produce pharmacologically distinct adrenomedullin receptors: a comparison of effects of adrenomedullin22-52, CGRP8-37 and BIBN4096BS.

    British journal of pharmacology·2003
    Same author

    Adaptive responses in human glioma cells assessed by clonogenic survival and DNA strand break analysis.

    International journal of radiation biology·2003
    Same author

    Effect of heat on phycoerythrin fluorescence: Influence of thermal exposure on the fluorescence emission of R-phycoerythrin.

    Biotechnology and bioengineering·2003
    Same author

    gamma-MSH increases intracellular cAMP accumulation and GnRH release in vitro and LH release in vivo.

    FEBS letters·2003
    Same journal

    Contribution of thebesian veins to the physiologic shunt in anesthetized man.

    Journal of applied physiology·2011
    Same journal

    Venous ultrasound catheter-tip technique for evaluation of arterial hemodynamics.

    Journal of applied physiology·1976
    Same journal

    Flow field and mass transport analysis in arteries with longitudinal ridges.

    Journal of applied physiology·1976
    Same journal

    Cardiorespiratory responses of sedentary college women as a function of training intensity.

    Journal of applied physiology·1976
    Same journal

    Genetic influence on normal variability of maximum expiratory flow-volume curves.

    Journal of applied physiology·1976
    Same journal

    Method to optimize high-pressure, multicomponent gas mixing.

    Journal of applied physiology·1976
    See all related articles

    Area of Science:

    • Physiology
    • Biomedical Engineering
    • Surgical Instrumentation

    Background:

    • Accurate control of arterial blood pressure is crucial for physiological research.
    • Manual methods for controlling arterial pressure in animal models are often technically challenging and imprecise.
    • Investigating the effects of transient or steady-state arterial pressure changes on organ function requires reliable pressure manipulation tools.

    Purpose of the Study:

    • To design and validate a servo-controlled aortic clamp for precise control of femoral arterial pressure in anesthetized rats.
    • To assess the accuracy, response time, and long-term stability of the developed aortic clamp system.
    • To provide a more convenient and accurate method for manipulating arterial pressure in experimental settings.

    Main Methods:

    Related Experiment Videos

    • A servo-controlled aortic clamp system was designed, including mechanical and electrical components.
    • The system was tested in anesthetized rats to control femoral arterial pressure by compressing the abdominal aorta.
    • System performance was evaluated based on pressure control accuracy, transient response, and long-term stability.

    Main Results:

    • The servo-controlled aortic clamp achieved long-term (30 min) mean pressure control within ±2 mmHg.
    • The system exhibited rapid response times, with step changes in pressure achieved in less than 2 seconds.
    • Short-term transients of ±3 mmHg were observed, indicating high precision.
    • The clamp system demonstrated superior accuracy and convenience compared to manual pressure control methods.

    Conclusions:

    • The developed servo-controlled aortic clamp offers accurate and convenient control of aortic pressure in rats.
    • This system minimizes technical difficulties associated with manual pressure manipulation.
    • The device facilitates research into the effects of arterial pressure variations on organ function, such as renal blood flow and glomerular filtration rate.