[Analysis of impulse response of rectangular ultrasound transducer]
1Department of Biomedical Engineering, Xi'an Jiaotong University, Xi'an 710049.
Abstract:
A new expression for calculating the impulse response of rectangular pistonlike transducers, proper to any points of acoustic field, is presented. The radiated field of ultrasound transducer can be analyzed using impulse response method, avoiding paraxial and farfield approximation. The velocity potential at an observation point is expressed as the convolution of the excitation velocity and the impulse response. A time-amplitude shift method for the far field impulse response of small aperture transducers is presented. The radiated field of rectangular transducers is calculated through the new method and the previous method and excellent agreement is observed.