Evaluation of the Dade Behring Dimension RxL clinical chemistry analyzer
Related Experiment Video
Updated: Oct 9, 2026

Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes
Published on: March 22, 2022
S Cuka1, S Dvornik, K Drazenović
1Clinical Hospital Center Rijeka, Department for Laboratory Diagnostics, Croatia. smiljana.cuka@ri.tel.hr
Abstract:
The performance of the Dade Behring Dimension RxL clinical chemistry analyzer was evaluated according to the guidelines of the European Committee for Clinical Laboratory Standards. The following analytes were tested: glucose, urea, creatinine, albumin, phosphorus, cholesterol, triglyceride, uric acid, magnesium, sodium, potassium, chloride, calcium, iron and total bilirubin. The Dade Behring Dimension RxL was compared with the Hitachi 704, Bayer RA-1000, Ektachem 250 and Chiron 865 depending on available tests on these analyzers. Coefficients of correlation showed high correlation between compared analyzers. Other performances (intra- and inter-assay variation, carry-over and interferences) of the analyzer were satisfactory.