NAL-NL1 procedure for fitting nonlinear hearing aids: characteristics and comparisons with other procedures
Journal of the American Academy of Audiology
|February 24, 2001
Summary
A new nonlinear hearing aid fitting procedure, NAL-NL1, aims to improve speech clarity while maintaining normal loudness. This method offers different gain and compression settings compared to other hearing aid fitting approaches.
Area of Science:
- Audiology
- Hearing Aid Technology
- Speech Perception
Background:
- Traditional hearing aid fitting procedures may not optimally balance speech intelligibility and loudness perception.
- The Speech Intelligibility Index (SII) is a key metric for predicting speech understanding, but requires modification for varying hearing loss degrees.
Purpose of the Study:
- To introduce and describe the National Acoustic Laboratories' nonlinear fitting procedure, version 1 (NAL-NL1).
- To compare NAL-NL1 fitting prescriptions with existing methods like DSL [i/o], FIG6, and IFF.
- To evaluate the impact of NAL-NL1 on gain and compression across different audiogram configurations.
Main Methods:
- Development of the NAL-NL1 procedure based on maximizing speech intelligibility (using a modified SII) and constraining loudness.
- Comparative analysis of NAL-NL1 prescriptions against NAL-Revised, Profound, DSL [i/o], FIG6, and IFF threshold procedures.
- Evaluation of prescribed gain and compression ratios for various audiogram shapes (flat, upward sloping, steeply sloping high-frequency).
Main Results:
- NAL-NL1 prescriptions closely resemble NAL-Revised, Profound for average speech input levels.
- NAL-NL1 generally prescribes lower low-frequency gain for flat/upward sloping audiograms and less high-frequency gain for steeply sloping losses compared to other methods.
- The NAL-NL1 procedure tends to recommend less overall compression than the compared fitting strategies.
Conclusions:
- The NAL-NL1 procedure offers a distinct approach to hearing aid fitting, potentially optimizing audibility and comfort.
- Significant differences exist between NAL-NL1 and other common hearing aid fitting protocols, particularly in gain and compression settings.
- Further research may be warranted to validate the clinical efficacy of NAL-NL1 across diverse patient populations.