[Tracheal tube damage by Harmonic Scalpel during tonsillectomy]
M Hayakawa1, Y Morimoto, O Kemmotsu
1Department of Anesthesiology and Critical Care Medicine, Hokkaido University School of Medicine, Sapporo 060-8638.
Abstract:
We experienced a case in which a tracheal tube was damaged by a Harmonic Scalpel during tonsillectomy. A 33-year-old man with chronic tonsillitis underwent tonsillectomy under general anesthesia. The trachea was intubated by Polar Preformed Tracheal Tube (Polar tube). About 45 minutes after the start of surgery, sudden massive air leakage in the trachea occurred. Surgery was accomplished under oral suctioning. After extubation, we found that an inflation cuff tube was cut off by the Harmonic Scalpel. The cause of damage was the position of an inflation cuff tube which had been in the dorsal side of the tracheal tube (near the surgical field) in the Polar tube. Anesthesiologists should be aware of the characteristics of the surgical equipment newly introduced.
Related Concept Videos
Trachea
Anatomical Features:
Location: About half of the trachea is situated in the neck, anterior to the esophagus, and extends from the larynx (at the level of the...
Tonsillitis II: Management
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Tracheostomy Care I: Pre-procedural Steps
Required Equipment
The equipment necessary for tracheostomy care includes:
Tracheostomy Suctioning II: Procedure