A four-column parallel chromatography system for isocratic or gradient LC/MS analyses
C K Van Pelt1, T N Corso, G A Schultz
1Advanced BioAnalytical Services, Inc., Ithaca, New York 14850, USA.
Analytical Chemistry
|February 24, 2001
Summary
This study introduces a novel parallel liquid chromatography/tandem mass spectrometry (LC/MS/MS) method. The technique enhances sample throughput for pharmacokinetic assays by analyzing multiple samples simultaneously, improving efficiency.
Area of Science:
- Analytical Chemistry
- Biochemistry
- Pharmacology
Background:
- Quantitative bioanalysis, particularly pharmacokinetic assays, demands high sample throughput.
- Conventional liquid chromatography/tandem mass spectrometry (LC/MS/MS) systems can be a bottleneck in high-throughput laboratories.
- Existing methods often require significant system modifications or specialized equipment.
Purpose of the Study:
- To present a novel, cost-effective parallel chromatography approach for LC/MS/MS analyses.
- To demonstrate the system's capability for enhancing sample throughput in quantitative applications.
- To validate the method's accuracy and precision using a real-world pharmacokinetic assay.
Main Methods:
- Modification of a conventional LC/MS system with three valves and four LC columns.
- Staggered injection of samples onto the four columns to enable continuous mass spectrometer analysis.
- Optimization of run time to slightly exceed the sum of desired peak widths.
- Application of both gradient and isocratic LC conditions.
- Analysis of atorvastatin, its metabolites, and deuterated internal standards in spiked human plasma.
Main Results:
- The parallel chromatography system achieved a throughput increase by a factor of 2.7 compared to conventional methods.
- Overall run time was reduced from 4.5 to 1.65 minutes for the specific analysis.
- Quality control samples exhibited relative standard deviation and accuracy within 8.1% and 9.6%, respectively.
- The system demonstrated robust performance under gradient elution conditions.
Conclusions:
- The developed parallel LC/MS/MS system offers a significant improvement in sample throughput for pharmacokinetic and quantitative assays.
- Modest system modifications allow for parallel analysis without compromising accuracy or precision.
- This approach provides a practical and efficient solution for high-throughput bioanalytical laboratories.