Accuracy of a noninvasive temporal artery thermometer for use in infants
1Division of Emergency Medicine, Children's Hospital, 300 Longwood Ave, Boston, MA 02115, USA. david.greenes@tch.harvard.edu
Objectives:
To assess the accuracy of a new noninvasive temporal artery (TA) thermometer in infants; to compare the accuracy of the TA thermometer with that of a tympanic thermometer, using rectal thermometry as the criterion standard; and to compare the tolerability of the TA thermometer with that of the tympanic and rectal thermometers.
Design:
Prospective evaluation of the accuracy of TA and tympanic thermometry, using rectal thermometry as the criterion standard.
Setting:
Emergency department of an urban pediatric hospital.
Subjects:
Convenience sample of 304 infants younger than 1 year presenting for care.
Main Outcome Measures:
Temperatures were measured using TA, tympanic, and rectal thermometers for all infants. Agreement between TA or tympanic and rectal temperatures was assessed. The sensitivity and specificity of TA or tympanic thermometers for detecting rectal fever were determined. Discomfort scores, using a standardized scale, were assessed by trained observers after each temperature measurement was made.
Results:
Linear regression analysis of the relation between TA and rectal temperatures yielded a model with a slope of 0.79 (vs a slope of 0.68 for tympanic vs rectal temperature; P =.02) and an r of 0.83 (vs r = 0.75 for tympanic vs rectal temperature; P<.001). Among 109 patients with a rectal temperature of 38 degrees C or higher, the TA thermometer had a sensitivity of 0.66 compared with the tympanic thermometer's sensitivity of 0.49 (P<.001). Discomfort scores with TA thermometry were significantly lower than with rectal thermometry (P =.007).
Conclusions:
The TA thermometer has limited sensitivity for detecting cases of rectal fever in infants. However, the TA thermometer is more accurate than the tympanic thermometer in infants, and it is better tolerated by infants than rectal thermometry.
Related Concept Videos
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...