Comparison of the coulometric and polarographic measurement of a high-Dk hydrogel
C F Morgan1, N A Brennan, L Alvord
1CIBA Vision Corporation, Atlanta, Georgia 30155-1518, USA. courtney.morgan@cibavision.novartis.com
Background:
International Organization for Standardization (ISO) document ISO 9913-1 details procedures for oxygen permeability coefficient (Dk) measurement of hydrogel contact lens materials up to 100 barrer. A draft ISO document 9913-2 describing the coulometric technique states that the recommended procedures are suitable for nonhydrogel materials. New generation hydrogel lens materials exceed 100 barrer, which is outside the scope of both documents.
Methods:
A range of rigid gas-permeable materials from Dr. W. Benjamin's standard repository was used as a baseline to assess the accuracy of the polarographic and the coulometric methods. We also measured the permeability of a new generation hydrogel, lotrafilcon A, to explore the suitability of the polarographic and coulometric techniques for high-Dk hydrogel lenses. We modified equipment for both methods that incorporate several improvements including front surface masking to eliminate "edge effect."
Results:
The coulometric technique provided results similar to those previously reported for the standard rigid gas-permeable materials and yielded relative standard errors typically <10%. The polarographic technique provided results similar to the accepted values for the standard rigid gas-permeable materials Dk <70 barrer only after the edge effect correction algorithm was applied. The polarographic results showed poor precision and significant, systematic differences from accepted values for the rigid gas-permeables with Dk > 70 barrer.
Conclusions:
The coulometric method is preferable for the measurement of contact lens materials with permeability >70 barrer. The coulometric method was successfully modified to include hydrogels.
Related Concept Videos
Voltammetric Techniques: Linear-Scan (E vs Time)
Coulometry: Overview
Controlled-Potential Coulometry: Electrolytic Methods
The chosen potential ensures...
Potentiometry: Membrane Electrodes
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...