Related Experiment Videos
[Intraoperative ventilatory obstruction by twisting of a reusable coiled expiratory breathing tube]
H Cuche1, J Bientz, J C Otteni
1Service d'anesthésie-réanimation, hôpital Edouard-Herriot, place d'Arsonval, 69437 Lyon, France.
Annales Francaises D'Anesthesie Et De Reanimation
|March 10, 2001
Abstract:
A case of intraoperative subtotal obstruction of a reusable coiled expiratory breathing tube is reported. Partial occlusion by twisting was made possible by detachment of the coil from the external face of the tube after multiple reprocessings with high drying temperatures. A technique for tube checking before reuse is described.