Related Experiment Video
Updated: Jun 30, 2026

Comprehensive & Cost Effective Laboratory Monitoring of HIV/AIDS: an African Role Model
Published on: November 1, 2010
Use of formalin-fixed, propidium iodide-stained human leukocytes as a standard for enumerating CD4+ T lymphocytes in
G M Harrison1, A J Bennett, M Moody
1Department of Medical Biochemistry and Immunology, University Hospital of Wales, Cardiff CF14 4XW, United Kingdom.
Abstract:
A new, inexpensive method is described that enables lymphocytes to be enumerated very precisely. Normal leukocytes were simultaneously stained and fixed with a propidium iodide-paraformaldehyde solution. The preparation obtained (CellBeads) was used as an internal standard for cell enumeration by flow cytometry and was stable at 4 degrees C for at least 60 days. Unlike synthetic beads, the CellBeads behaved similarly to normal cells during red blood cell lysis and cell washing procedures. When known numbers of CellBeads were added to whole-blood samples and the numbers of CellBeads and lymphocytes were determined, highly reproducible and accurate enumerations were obtained-far more so than when synthetic beads were used. This inexpensive method is suitable for routine use.