[Usefulness of a VDD defibrillation electrode in recording atrial electrograms during atrial flutter and atrial
M Niehaus1, G Thuermer, S Thamasett
1Abteilung für Kardiologie und Angiologie, Medizinische Hochschule Hannover. M.Niehaus@T-Online.de
Background:
Monitoring of atrial signals improves the accuracy in identifying supraventricular tachyarrhythmias to prevent inappropriate therapies in patients with implantable cardioverter-defibrillators (ICD). Since complications due to the additional atrial lead were found in dual chamber ICD systems with 2 leads, we designed a single-pass VDD-lead for use with dual chamber ICDs.
Patients And Methods:
After promising animal experiments in a German multicenter study a prototype VDD lead (single-coil defibrillation electrode with 2 additional fractally coated rings for bipolar sensing in the atrium) was temporarily used in 20 patients. Atrial and ventricular signals were recorded during sinus rhythm, atrial flutter, atrial fibrillation and ventricular tachycardia or ventricular fibrillation. Terminations of ventricular arrhythmias were performed by internal DC shock.
Results:
The implantation of the electrode was successful in 18 of 20 patients. Mean atrial pacing threshold was 2.45 +/- 0.9 V/0.5 ms, mean atrial impedance was 215 +/- 31 Ohm. Atrial amplitudes were greater during sinus rhythm (2.7 +/- 1.6 mV) than during atrial flutter (1.36 +/- 0.28 mV, p < 0.05) or atrial fibrillation (0.92 +/- 0.29 mV, p < 0.01). During ventricular fibrillation atrial "sinus"-signals had significantly (p < 0.01) lower amplitudes than during sinus rhythm. Mean ventricular sensing was 13.3 +/- 7.9 mV, mean ventricular impedance was 577 +/- 64 Ohm. Defibrillation was successful with 20 J shock. 99.6% of P waves could be detected in sinus rhythm and 85 +/- 9.9% of flutter waves during atrial flutter. During atrial fibrillation 55% of atrial signals could be detected without modification of the signal amplifier.
Conclusions:
A new designed VDD dual chamber electrode provides stable detection of atrial and ventricular signals during sinus rhythm and atrial flutter. For reliable detection of atrial fibrillation modifications of the signal amplifier are necessary.
More Related Videos
Related Concept Videos
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and the T...
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...
Dysrhythmias V: Evaluating Dysrhythmias
Cardiopulmonary Resuscitation III: AED Use
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias