[Imaging quality of a digital Konica Regius 336 system for thoracic radiology]
A Ostinelli1, M Frigerio, A F Monti
1Servizio di Fisica Sanitaria, Azienda ospedaliera S. Anna, Via Napoleona 60, 22100 Como CO. fisan.como@tiscalinet.it
Introduction:
Digital radiographic systems permit to optimize execution, depiction and storage of radiological images. Since a Regius 336 digital system (Konica Corp., Tokyo, Japan) devoted to chest radiography was recently installed in the Radiology Department of S. Anna Hospital in Como, Italy, we investigated its performance relative to image quality.
Material And Methods:
Konica Regius 336 is a computed radiography system made of a phosphorescence detector plate which is scanned with an infrared semiconductor laser beam. The radiographic image obtained from the detector is subjected to image processing, which allows a stable output and the nonlinear curve typical of conventional radiographic systems. Image quality was assessed based on the following parameters: dose, contrast, noise, and spatial resolution. As reference, we assessed the same parameters on a Cronex 88 analogic chest-changer (DuPont Pharma, North Billerica, Mass, USA).
Results:
The Regius 336 air kerma values were always higher than the analogic ones (about 10%), both with and without a chest phantom; noise was also greater than in analogic images, sometimes even doubled. The optical densities of a step wedge and the spatial resolution of the digital chest-changer are independent of the X-ray tube voltage consequent to broader optical latitude. Inversely, the analogic images of the wedges show great optical density variability as a function of the X-ray tube voltage (in a range of 2). The modulation transfer functions of the two systems have the same trend.
Discussion And Conclusions:
The performance of the Konica Regius 336 is nearly equivalent to that of an analogic system. The main advantages of the digital system are a standard output, lower consumption of radiographic films, higher productiveness and better image quality standard level.
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies III: Computed Tomography
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Radiological Investigation I: X-ray and CT
X-ray Imaging
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...