MR evaluation of ventricular function: true fast imaging with steady-state precession versus fast low-angle shot cine
J Barkhausen1, S G Ruehm, M Goyen
1Department of Diagnostic Radiology, University Hospital Essen, Hufelandstrasse 55, D-45122 Essen, Germany. joerg.barkhausen@uni-essen.de
Abstract:
Short- and long-axis cine magnetic resonance (MR) images were obtained with a standard fast low-angle shot, or FLASH, sequence and a first-generation true fast imaging with steady-state precession (FISP) sequence on a 1.5-T MR imager. Contrast-to-noise ratios and volumetric left ventricular measurements were compared for manual and automatic segmentation. True FISP images were associated with significantly (P<.01) higher contrast-to-noise ratios and allowed better detection of the endocardial border. True FISP images were provided with short acquisition times and excellent contrast between the myocardium and the ventricular lumen.