Infusion pump performance with vertical displacement: effect of syringe pump and assembly type
T A Neff1, J E Fischer, G Schulz
1Department of Intensive Care and Neonatology, University Children's Hospital, Zurich, Switzerland.
Objective:
To evaluate the effect of different infusion pump models on continuity of drug delivery during vertical displacement of syringe pumps.
Design:
Zero-drug delivery time (ZDDT), retrograde aspiration volume, and infusion bolus were recorded using the same syringe in three different models of syringe pump after lowering and elevating the pump. Compliance of each infusion assembly was measured using the occlusion release technique at 38 mmHg.
Results:
Lowering the pump by 50 cm at an infusion rate of 1 ml/h resulted in ZDDT values ranging from 2.78 +/- 0.29 to 5.99 +/- 1.09 min. Elevating the syringe pump to its original position caused infusion boluses between 44.1 +/- 3.2 and 77.1 +/- 5.1 microl. The results demonstrated that there are large differences between syringe pump models (F = 66.8, df = 2/33, p < 0.0001) and between pumps of the same model (F = 21.3, df = 1/34, p < 0.0001). A similar pattern was found in retrograde aspiration volume and infusion bolus.
Conclusion:
All tested pumps led to clinically relevant flow irregularities during vertical displacement of the syringe pump. Thus, vertical displacement of any syringe pump connected to an infusion line delivering highly potent drugs at low infusion rates should be avoided. The variability across syringe pumps indicates that syringe pump design remains an area of potential further improvement for reducing the risk of adverse patient events.
Related Concept Videos
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...
Parenteral Drug Delivery Systems: Injectables, Implants, and Infusion Devices
Two-Compartment Open Model: IV Infusion
The model illustrates the decrease in plasma drug concentration from the central compartment with a specific equation. It shows that under steady-state conditions, the drug's input rate...
Compartment Models: Single-Compartment Model
Application of the Energy Equation
Drug Accumulation During Multiple Dosing: Intermittent IV Infusions