MR imaging of vascular stents: effects of susceptibility, flow, and radiofrequency eddy currents
L W Bartels1, H F Smits, C J Bakker
1Department of Radiology, Image Sciences Institute, University Medical Center Utrecht, Heidelberglaan, The Netherlands. w.bartels@azu.nl
Purpose:
The purpose of this in vitro study was to examine the various sources of artifacts in magnetic resonance (MR) imaging and angiography of vascular stents.
Materials And Methods:
Five low-artifact stents-Wallstent (cobalt alloy), Memotherm (nitinol), Perflex (stainless steel), Passager (tantalum), and Smart (nitinol)-were imaged in a vascular flow phantom, consisting of a thin-walled cellulose vessel model connected to a pump system. The echo time and the angulation of the stents with respect to the direction of the main magnetic field were varied. Spin echo and gradient echo images as well as three-dimensional MR angiograms were obtained to study the effects of flow, magnetic susceptibility, and radiofrequency-induced eddy currents.
Results:
Susceptibility artifacts were restricted to the stents' direct environment and were mildest at short echo times and with the stents aligned with the main magnetic field. Nitinol stents showed less artifacts than steel stents did. Radiofrequency artifacts obscuring the stent lumen and flow-related lumen displacement were seen in all stents. The extent to which these occurred depended on strut geometry and orientation.
Conclusions:
For low-artifact stents, the material the stent is made of is not the only important factor in the process of artifact formation. Susceptibility artifacts, radiofrequency eddy currents and flow-related artifacts all contribute to the image distortion, and are dependent on the geometry and orientation of the struts and on the orientation of the stent in the main magnetic field.
Related Concept Videos
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...
Magnetic Resonance Imaging