Related Experiment Video
Updated: Oct 8, 2026

Development of a Colloidal Gold-based Immunochromatographic Test Strip for Detection of Cetacean Myoglobin
Published on: July 13, 2016
Use of an immunoassay as a means to detect polychlorinated biphenyls in animal fat
S Jaborek-Hugo1, C von Holst, R Allen
1European Commission, DG Joint Research Centre, Institute for Health and Consumer Protection, Food Products and Consumer Goods Unit, Ispra, Italy.
Abstract:
In this paper we describe the validation and utilization of the Hybrinzme DELFIA immunoassay for detecting polychlorinated biphenyls (PCBs) in animal fat. The immunoassay incorporates a rapid sample processing protocol that has been optimized to detect concentrations of PCBs in animal fat at or above 200 ng/g. The limit of detection of the assay was less than 50 ng/g. When the action level for the DELFIA assay was set at 100 ng/g to detect samples having 200 ng/g or more PCBs, the false negative rate was estimated at 0.2%. The false positive rate depends on the concentration of PCBs in the sample population and was estimated at 4% and 7% for samples having 10 ng/g and 20 ng/g PCBs, respectively. By utilizing the DELFIA assay system to select for positive samples and GC/MS analysis for confirmation analysis, an accurate and cost-effective programme has been established for the high throughput detection of PCBs in animal products.