Direct Pressure Monitoring Accurately Predicts Pulmonary Vein Occlusion During Cryoballoon Ablation
Published on: February 26, 2013
[Accuracy of pressure measurements obtained with a new rectal balloon catheter]
L Le Normand1, P Glemain, J P Cordonnier
1Clinique Urologique, CHU Nantes, France. llenormand@chu-nantes.fr
Unlabelled:
The accuracy of measurements performed by a balloon catheter used to record abdominal pressure during urodynamic investigations was verified on a test bench.
Objective:
To study the accuracy and precision of the pressure measurements obtained with a new rectal balloon catheter (ref. 95018 Laboratoires Vermed), designed to record abdominal pressure during urodynamic assessments. The clinical value of this catheter, using air for pressure transmission, is its simplicity, as there is no contamination of the perfusion circuit or transducer, no purging of the circuit, and artefacts related to movements of the tubing are eliminated.
Method:
The catheter was placed in a pressure chamber fitted with a precise, calibrated regulation system allowing programmed pressure variations from 10 to 150 cmH2O. Pressures recorded by the test catheter were compared to reference pressures applied to the chamber. The frequency of acquisition of pressure measurements was 100 Hertz and the resolution was 10 Hertz. This model was used to study the accuracy of pressure measurements and the response times of the catheter. Measurements were performed with a volume of 2 ml of air introduced into the catheter (volume recommended by the manufacturer), and the optimal volume was investigated by inflating the catheter until the best result was obtained. The evaluation was based on calculation of the mean difference observed between the two measurements and the scatter of the differences observed.
Results:
When the catheter was filled with 2 ml of air, pressures measured by the catheter were overestimated an average of 1.1 cmH2O (standard deviation = 1), and 95% of the differences between the two measurements were within +/- 2.15 cmH2O. The optimal air volume was found to be 1.5 ml. With this air volume, no significant difference was observed between the two measurements. The mean observed difference was 0.2 cmH2O (SD = 1.2), which means that 95% of the differences were situated within the range of +/- 2.35 cmH2O.
Conclusion:
The pressure recording method with this new catheter is validated in terms of physical parameters.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Errors occurring during blood pressure monitoring
Several factors...
Assessment of blood pressure in brachial artery(two-step method)
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure