Reliability of manometric eustachian tube function tests in children
N van Heerbeek1, K J Ingels, A F Snik
1Department of Otorhinolaryngology, University Medical Center St. Radboud, Nijmegen, The Netherlands.
Summary
Manometric eustachian tube function tests show good reproducibility in children with ventilation tubes. A single measurement or the mean of three is sufficient for assessing tubal function.
Area of Science:
- Otolaryngology
- Pediatric Audiology
- Medical Engineering
Background:
- Persistent otitis media with effusion (OME) in children often necessitates ventilation tube insertion.
- Accurate assessment of eustachian tube function is crucial for managing OME and guiding treatment.
- Manometric tests offer a quantitative method to evaluate eustachian tube patency.
Purpose of the Study:
- To evaluate the reliability and reproducibility of manometric eustachian tube function tests in pediatric patients with ventilation tubes.
- To determine the optimal number of measurements for accurate assessment of tubal function.
Main Methods:
- Manometric eustachian tube function tests, including opening pressure (Po) and closing pressure (Pc), were performed repeatedly within a single session.
- Ninety-nine children with ventilation tubes in situ due to persistent OME participated.
- Tests were conducted at secondary and tertiary referral hospitals.
Main Results:
- Analysis revealed a statistically significant decrease in both Po and Pc with repeated measurements (p < 0.001).
- The reduction in Po was more pronounced than in Pc upon repeated testing.
- Excellent agreement (99%) was observed between the first and second pressure equilibration tests.
Conclusions:
- The categorized results of the pressure equilibration test demonstrate good reproducibility in children with ventilation tubes.
- The forced-response test showed lower reproducibility and a downward shift with repeated measurements.
- A single measurement or the mean of the first three Po and Pc measurements is adequate for determining tubal function.