Reliability of manometric eustachian tube function tests in children

N van Heerbeek1, K J Ingels, A F Snik

  • 1Department of Otorhinolaryngology, University Medical Center St. Radboud, Nijmegen, The Netherlands.

Insights

Manometric eustachian tube function tests show good reproducibility in children with ventilation tubes. A single measurement or the mean of three is sufficient for assessing tubal function.

Area of Science:

  • Otolaryngology
  • Pediatric Audiology
  • Medical Engineering

Background:

  • Persistent otitis media with effusion (OME) in children often necessitates ventilation tube insertion.
  • Accurate assessment of eustachian tube function is crucial for managing OME and guiding treatment.
  • Manometric tests offer a quantitative method to evaluate eustachian tube patency.

Purpose of the Study:

  • To evaluate the reliability and reproducibility of manometric eustachian tube function tests in pediatric patients with ventilation tubes.
  • To determine the optimal number of measurements for accurate assessment of tubal function.

Main Methods:

  • Manometric eustachian tube function tests, including opening pressure (Po) and closing pressure (Pc), were performed repeatedly within a single session.
  • Ninety-nine children with ventilation tubes in situ due to persistent OME participated.
  • Tests were conducted at secondary and tertiary referral hospitals.

Main Results:

  • Analysis revealed a statistically significant decrease in both Po and Pc with repeated measurements (p < 0.001).
  • The reduction in Po was more pronounced than in Pc upon repeated testing.
  • Excellent agreement (99%) was observed between the first and second pressure equilibration tests.

Conclusions:

  • The categorized results of the pressure equilibration test demonstrate good reproducibility in children with ventilation tubes.
  • The forced-response test showed lower reproducibility and a downward shift with repeated measurements.
  • A single measurement or the mean of the first three Po and Pc measurements is adequate for determining tubal function.
Abstract