Effect of inner cannula removal on the work of breathing imposed by tracheostomy tubes: a bench study
T Cowan1, T B Op't Holt, C Gegenheimer
1Department of Cardiorespiratory Care, University of South Alabama, Mobile, USA. Mucomist@email.com
Background:
Tracheotomy has been used to assist in weaning patients from mechanical ventilation. Some patients fail to be weaned from the ventilator despite tracheostomy. We hypothesized that removing the inner cannula from the tracheostomy tube would decrease the tube's imposed work of breathing (WOB(IMP)).
Methods:
The hypothesis was tested using a lung model, by measuring the change in WOB(IMP) when the inner cannula was removed. A mechanical lung model was developed using a test lung to simulate a spontaneously breathing patient. WOB(IMP) was measured with a commercially available lung mechanics monitor. Shiley size 6, 8, and 10 nonfenestrated tracheostomy tubes were tested with the inner cannula in and out. Breathing conditions were simulated using tidal volumes (V(T)) of 300 and 500 mL matched with breathing frequencies of 12, 24, and 32 breaths per minute, by using a ventilator to simulate spontaneous breathing through one side of the test lung.
Results:
Under all the tested breathing conditions, WOB(IMP) for each of the 3 tracheostomy tubes was significantly reduced (p < 0.05) when the inner cannula was removed. Also, as simulated spontaneous inspiratory flow demand increased (ie, as V(T) and/or frequency were increased), WOB(IMP) also increased, and vice versa. With the cannula removed, WOB(IMP) was not significantly different between the size 6 and 8 tubes nor between the size 8 and 10 tubes when V(T) was 300 mL and frequency was 12 breaths per minute.
Conclusions:
There was a significant decrease in WOB(IMP) with each tube when the inner cannula was removed. WOB(IMP) increased with an increase in inspiratory flow demand (ie, increase in V(T) and/or frequency), as well as when tube size decreased. In weaning a tracheostomized patient from mechanical ventilation, increasing the internal diameter of the tube by removing the inner cannula may be beneficial. Further study is needed to determine if these findings are clinically important.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Tracheostomy Care I: Pre-procedural Steps
Required Equipment
The equipment necessary for tracheostomy care includes:
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...
Tracheostomy Care II: Procedure
Step 1: Perform hand hygiene, and put on personal protective equipment: gown, gloves, mask and...