Ancillary tools in pacemaker and defibrillator lead extraction using a novel lead removal system
A S Manolis1, T N Maounis, V Vassilikos
1Cardiology Section, Patras University, Patras, Greece. asm@otenet.gr
Pacing and Clinical Electrophysiology : PACE
|April 20, 2001
Summary
Percutaneous lead extraction using the VasoExtor system is highly successful, but ancillary tools are often needed to enhance outcomes for pacemaker and implantable cardioverter-defibrillator (ICD) leads. This study highlights the necessity of supplementary devices in over a quarter of cases for complete lead removal.
Area of Science:
- Cardiology
- Medical Devices
- Interventional Procedures
Background:
- Pacemaker and implantable cardioverter-defibrillator (ICD) lead extraction is crucial for managing complications like infection or malfunction.
- The VasoExtor system offers a preliminary successful approach for percutaneous lead removal.
- Experience indicates the need for additional tools to optimize extraction success rates.
Purpose of the Study:
- To evaluate the efficacy of the VasoExtor system with various locking stylets (S, K, Magic) for pacemaker and ICD lead extraction.
- To determine the necessity and types of ancillary tools required to enhance lead extraction success.
- To assess the overall success and complication rates associated with percutaneous lead extraction.
Main Methods:
- A retrospective analysis of 34 patients undergoing extraction of 51 leads (48 pacemaker, 3 ICD) over 3 years.
- Utilized standard (S, K) and newer (Magic) locking stylets with the VasoExtor system.
- Documented the use of ancillary tools, including noose catheters, pigtail catheters, bioptomes, and Amplatz catheters, when standard methods were insufficient.
Main Results:
- Lead extraction was successful in 97% of patients (33/34) and 98% of leads (50/51) without major complications.
- Ancillary tools were required in 26.5% (9/34) of patients to facilitate extraction.
- Specific challenges included inability to lock leads and retrieval of lead fragments, necessitating combined or alternative approaches.
Conclusions:
- Percutaneous lead extraction using the VasoExtor system with locking stylets is highly effective.
- A significant proportion of cases require ancillary tools to achieve complete lead removal.
- The integration of various tools enhances the success rate of complex pacemaker and ICD lead extractions.