Related Experiment Videos
Effect of round-edged acrylic intraocular lenses on preventing posterior capsule opacification
Purpose:
To clarify the extent to which the adhesiveness of an acrylic material influences the formation of posterior capsule opacification (PCO).
Setting:
Jinshikai Medical Foundation, Nishi Eye Hospital, Osaka, Japan.
Methods:
Two types of AcrySof(R) intraocular lenses (IOLs) were prepared: round edged and tumbled. The AcrySof with round edges was implanted in 1 eye in a group of 4 rabbits and the tumbled IOL, in 1 eye in a group of 5 rabbits. In both groups, the contralateral eye received a conventional AcrySof with sharp optic edges. A histopathological examination was performed 3 weeks after surgery.
Results:
With the round-edged AcrySof IOL, no capsular bend formed at the optic edge and abundant lens epithelial cells (LECs) migrated posteriorly. With the sharp-edged AcrySof lens, a sharp capsular bend formed and LEC migration was significantly inhibited. In eyes with a tumbled IOL, a capsular bend was created, but it was less marked than that created by the sharp-edged lens and there was slightly more LEC migration posteriorly.
Conclusions:
The AcrySof IOL lost its preventive effect on PCO when the optic was rounded. The effect of the AcrySof lens in preventing PCO is mainly a result of its rectangular, sharp-edged optic design. The acrylic material may play a complementary role by helping create a sharp capsular bend. Capsular bend formation is the key to the PCO preventive effect of an IOL.