Determination of albumin and myoglobin in dialysate and ultrafiltrate samples by high-performance size-exclusion chromatography
Related Experiment Video
Updated: Jul 29, 2026

Protein Digestion, Ultrafiltration, and Size Exclusion Chromatography to Optimize the Isolation of Exosomes from Human Blood Plasma and Serum
Published on: April 13, 2018
Abstract:
A high-performance size-exclusion chromatographic method was developed, validated and implemented for simultaneous and quantitative determination of albumin and myoglobin along with inulin, vancomycin and creatinine in dialysate and ultrafiltrate samples from in vitro hemodialysis experiments. The experimental parameters including mobile phase pH, ionic strength, detection wavelength, flow-rate, injection volume were first optimized for the determination of albumin, myoglobin, inulin, vancomycin and creatinine. The peak height ratio and detection limits of the proteins were then comparatively studied at 210, 254 and 280 nm by UV and diode array detection. The method was further validated by evaluating the linearity, precision and accuracy of the proteins. The assay was finally implemented to the simultaneous and quantitative determination of the proteins in dialysate and ultrafiltrate samples.
Related Concept Videos
Types Of Column Chromatography
Gel Filtration Chromatography
When the...
Dialysis
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
Ion-Exchange Chromatography
Size-Exclusion Chromatography
Silica particles offer advantages such as rigidity,...
Affinity Chromatography