Evaluation of an automated liquid-handling system (Tecan Genesis RSP 100) in the Abbott LCx assay for Chlamydia
1Department of Laboratory Medicine and Pathology, Hennepin County Medical Center, 701 Park Ave., Minneapolis, MN 55415, USA.
Abstract:
The present study investigated the feasibility of automating the specimen-pipetting component of sample preparation in the LCx Chlamydia assay (LCx-CT assay; Abbott Laboratories, Chicago, Ill.) by using a commercially available liquid-handling system (Tecan Genesis RSP100; Tecan Inc., Research Triangle Park, N.C.). The Tecan instrument proved to be comparable in both precision and accuracy to a manual multipipettor (Eppendorf model 4850; Eppendorf Scientific, Westbury, N.Y.). The Tecan instrument was extensively checked for evidence of specimen-to-specimen transfer, and no level of contamination sufficient to generate a signal above the background in the LCx-CT assay was detected. Finally, pipetting speed was significantly improved by using the Tecan instrument. A mean time of 2.5 min was required to pipette a complete LCx-CT assay carousel (20 samples and 4 controls) with the Tecan instrument, whereas 8.4 min was required to pipette a comparable number of samples manually (P < 0.001).